Description
HI 2300 performs EC, TDS (Total Dissolved Solids), percentage of NaCl and temperature measurements. The auto-ranging feature for EC and TDS measurements automatically sets the resolution suitable to the tested sample.
All measurements can be temperature compensated at 20 or 25°C and the compensation coefficient is selectable by the user. The automatic temperature compensation can also be disabled for measuring the actual conductivity value. The stability indicator on the LCD ensures accuracy.
Conductivity readings are performed by applying an alternate current to the 4-ring probe which creates a variable voltage directly proportional to conductivity.
The GLP feature allows users to store and recall data about system status.
Specifications
| Range | EC | 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm (actual EC) |
| Range | TDS | 0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L (ppt); 15.0 to 100.0 g/L (ppt); up to 400.0 g/L (actual TDS), with 0.80 conversion factor |
| Range | NaCl | 0.0 to 400.0% |
| Range | Temperature | -9.9 to 120.0°C |
| Resolution | EC | 0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm |
| Resolution | TDS | 0.01 mg/L; 0.1 mg/L; 1 mg/L; 0.01 g/L; 0.1 g/L |
| Resolution | NaCl | 0.1% |
| Resolution | Temperature | 0.1°C |
| Accuracy | EC | ±1% of reading ±(0.05 µS/cm or 1 digit) |
| Accuracy | TDS | ±1% of reading ±(0.03 mg/L or 1 digit) |
| Accuracy | NaCl | ±1% of reading |
| Accuracy | Temperature | ±0.4°C |
| Calibration | EC | automatic, 1 point with 6 memorized values (84, 1413, 5000, 12880, 80000, 111800 µS/cm) |
| Calibration | NaCl | 1 point, with HI 7037 calibration solution |
| Calibration | Temperature | 2 point, at 0 to 50°C (32 to 122°F) |
| Temperature Compensation | | Automatic or manual, 0 to 60°C (32 to 140°F) |
| Temperature Coefficient | | selectable from 0.00 to 6.00%/°C (EC and TDS only) |
| TDS Conversion Factor | | selectable from 0.40 to 0.80 (default value 0.50) |
| Probe | | HI 76310, 4-ring platinum, internal temperature sensor |
| Auto Shut-off | | after 5 minutes of non-use (can be disabled) |
| Power Supply | | 12 Vdc adapter (included) |
| Environment | | 0 to 50°C (32 to 122°F); RH max 95% |
| Dimensions | | 240 x 182 x 74 mm (9.4 x 7.2 x 2.9") |
| Weight | | 1.1 kg (2.4 lb.) |
Accessories
To View or Buy Any Accessory, Search For Item By Name (Click and Paste).
| Electrodes |
| HI 76310 | 4-Ring Potentiometric Probes |
| Solutions |
| HI 7030L | 12880 µS/cm (µmho/cm) 1 x 500 mL bottle |
| HI 7031L | 1413 µS/cm (µmho/cm) 1 x 500 mL bottle |
| HI 7033L | 84 µS/cm (µmho/cm) 1 x 500 mL bottle |
| HI 7034L | 80000 µS/cm (µmho/cm) 1 x 500 mL bottle |
| HI 7035L | 111800 µS/cm (µmho/cm) 1 x 500 mL bottle |
| HI 7039L | 5000 µS/cm (µmho/cm) 1 x 500 mL bottle |
| HI 7037L | Calibration solution for % Readings (100% NaCl) 1 x 500 mL bottle |
| Accessories |
| HI 710005 | Voltage adapter from 115V to 12 Vdc (USA plug) |
| HI 710006 | Voltage adapter from 230V to 12 Vdc (European plug) |
| HI 920010 | Cable for PC Connection, 9 to 9-pin |
| HI 92000 | Windows® Compatible Software for Logging Instruments, Photometers and Turbidity Meters |
Order Info
HI 2300 is supplied with HI 76310 probe, 12 Vdc power adapter and instructions.
Manufacturer: N/A
SKU: HI2300