Hanna HI9813-6   Waterproof pH/EC/TDS/ Temperature Meter with pH Electrode Diagnostic
 

Hanna HI9813-6 Waterproof pH/EC/TDS/ Temperature Meter with pH Electrode Diagnostic

Price: $207.00
Quantity:
Detailed Description
Specifications

Description:

pH electrodes are commonly designed to produce and output signal equal to 0 mV when immersed in a solution with pH 7. But a constant output value at pH 7 occurs as a result of certain adverse events, such as:

• A broken bulb in the pH electrode
• A short circuit in the cable or connector
• Magnetic force leakage in the cable or connector (for example, in case of excessive humidity)

In these situations, the instrument displays a false reading at pH 7 and the user may not be aware that corrective measures are necessary; particularly if the user is anticipating a pH neutral reading. To avoid this problem, HANNA® developed F.S.T. technology for HI 9813-6.

With this special security system, a system failure will force a fixed value close to pH 4, rather than pH 7, and render the problem clearer to the user.

For added confidence in your measurements, the meter is equipped with two LEDs and a "CHECK" button. While pressing this button, upon immersing the electrode in the Cal-Check solution HI 50021P, one of the LEDs will light up to indicate the status of the meter. The green LED ("OK") indicates the probe is working correctly, while the red LED ("C&C" Clean & Cal) indicates the electrode needs attention.

Specifications:

Range

pH

0.0 to 14.0 pH

Range

EC

0.00 to 4.00 mS/cm

Range

TDS

0 to 1999 ppm (mg/L)

Range

Temperature

0.0 to 60.0°C

Resolution

pH

0.1 pH

Resolution

EC

0.01 mS/cm

Resolution

TDS

1 ppm (mg/L)

Resolution

Temperature

0.1°C

Accuracy

pH

±0.1 pH

Accuracy

EC

±2% F.S.

Accuracy

TDS

±2% F.S.

Accuracy

Temperature

±0.5°C

TDS Conversion Factor

 

0.56 to 0.72 ppm = 1 µS/cm

Calibration

 

manual at 1 point (all parameters except temperature)

Temperature Compensation

 

automatic, 0 to 50°C (32 to 122°F) with ß = 2%/°C (EC/TDS only)

Probe

 

HI 1285-6 pH/EC/TDS/T,
with Clean & Cal, 1 m (3.3') cable (included)

Battery Type / Life

 

1 x 9V / approx. 150 hours of continuous use

Environment

 

0 to 50°C (32 to 122°F); RH max 100%

Dimensions

 

165 x 75 x 45 mm (6.9 x 2.9 x1.8")

Weight

 

180 g (6.3 oz.)


Accessories:

Electrodes

Click To View or Buy

Special Applications Specific Electrodes for Specific Analysis

Solutions

 

Click To View or Buy

Electrode Storage Solution
1 x 500 mL bottle

Click To View or Buy

General Purpose
1 x 500 mL bottle

Click To View or Buy

Combination pH/Conductivity Buffer Kit
1413 µS/cm & 7.01 pH

Click To View or Buy

Combination pH/TDS Buffer Kit
1500 ppm & 7.01 pH

Click to View or Buy

Check Solution,25 x 20 mL Sachets

Accessories

 

Click To View or Buy

Shockproof Boot

Click To View or Buy

Shockproof Boot

Click To View or Buy

Protective Case for HI 8424NEW


Order Info:
HI 9813-6 is supplied with pH/EC/TDS/T HI 1285-6 probe with 1 m (3.3') cable, pH 7, 1.41 mS/cm and 1500 ppm calibration solution (20 mL each), cleaning solution (2 pcs), HI 50021 test solution (2 pcs), blue protective case, battery and instructions.

Downloads:

Manual: (Download)
Literature: (Download)

Return to Catalog